What is Sensitivity Drift?

Published in Measurement Error 6 mins read

Sensitivity drift, also known as scale factor drift, defines the amount by which an instrument's sensitivity of measurement varies as ambient conditions change. This phenomenon is a critical consideration in metrology and instrument design, as it directly impacts the accuracy and reliability of measurements over time and across different operating environments.

Essentially, it means that for a given input, the instrument's output might change not because the input itself has changed, but because the instrument's response rate to that input has shifted. For example, if a pressure sensor is designed to output 10 millivolts (mV) per 1 Pound per Square Inch (PSI) of pressure, sensitivity drift means that under different temperatures, it might output 9.8 mV/PSI or 10.2 mV/PSI for the same pressure change.

Understanding the Core Concepts

To fully grasp sensitivity drift, it's helpful to break down its key components:

  • Sensitivity: This refers to the ratio of an instrument's output change to a corresponding input change. It determines how much the output signal varies for each unit of change in the measured quantity. For instance, a sensor with high sensitivity produces a larger output change for a small input change.
  • Drift: In the context of instrumentation, drift signifies a gradual, unintended change in an instrument's output or characteristics over time or due to varying conditions, even when the measured input remains constant.
  • Ambient Conditions: These are the environmental factors surrounding the instrument, such as:
    • Temperature: The most common cause of drift.
    • Humidity: Moisture content in the air.
    • Pressure: Barometric pressure fluctuations.
    • Vibration and Shock: Mechanical stresses.
    • Electromagnetic Interference (EMI): External electrical or magnetic fields.
    • Aging: Long-term material changes.

Why Sensitivity Drift Matters

The primary implication of sensitivity drift is a loss of measurement accuracy. Unlike zero drift (or offset drift), which causes a constant error regardless of the input value, sensitivity drift introduces an error that is proportional to the measured value. This means the error becomes more significant as the measured quantity increases.

Consequences of Unmanaged Sensitivity Drift:

  • Inaccurate Readings: Leading to incorrect decisions in critical applications.
  • Reduced Reliability: Instruments may not perform consistently.
  • Increased Maintenance Costs: Due to frequent recalibration requirements.
  • Product Quality Issues: In manufacturing, drift can lead to products outside specifications.
  • Safety Concerns: In applications like medical devices or aerospace, inaccuracies can be dangerous.

Common Causes of Sensitivity Drift

Various factors contribute to an instrument's sensitivity drift:

  • Temperature Variations:
    • Material Expansion/Contraction: Changes in the physical dimensions of sensor components can alter their mechanical and electrical properties.
    • Electrical Resistance: Resistors and semiconductor components change resistance with temperature, affecting circuit sensitivity.
    • Thermoelectric Effects: Dissimilar materials in contact can generate small voltages (Seebeck effect).
  • Aging and Material Degradation:
    • Component Wear: Gradual breakdown of mechanical parts.
    • Chemical Changes: Oxidation, corrosion, or contamination affecting sensor elements.
    • Stress Relaxation: Changes in material properties under prolonged stress.
  • Humidity:
    • Moisture Absorption: Can alter the dielectric properties of insulating materials or affect resistive elements.
    • Corrosion: Increased moisture accelerates the corrosion of metal components.
  • Mechanical Stress:
    • Vibration: Constant vibrations can loosen connections or induce micro-cracks.
    • Shock: Sudden impacts can permanently alter sensor characteristics.
  • Power Supply Fluctuations:
    • Instability in the power supplied to an instrument's electronic components can directly affect its output gain and, consequently, its sensitivity.

Practical Examples

  • Pressure Sensors: A piezoresistive pressure sensor's output (voltage per unit pressure) might decrease as the ambient temperature rises, leading to under-readings at higher temperatures for the same actual pressure.
  • Load Cells (Weight Sensors): The strain gauges within a load cell can exhibit changes in their gauge factor (sensitivity to strain) with temperature, meaning a given weight might produce a different electrical signal depending on the temperature.
  • Thermocouples: While primarily temperature sensors, the output voltage per degree Celsius can slightly vary with aging or contamination of the junction, affecting overall sensitivity.
  • Analog-to-Digital Converters (ADCs): The reference voltage used by an ADC can drift with temperature, influencing the scaling of the digital output relative to the analog input.

Mitigating and Managing Sensitivity Drift

Addressing sensitivity drift is crucial for maintaining measurement accuracy. Several strategies are employed:

  • Temperature Compensation:
    • Passive Compensation: Using resistors or thermistors with specific temperature coefficients to counteract the sensor's drift.
    • Active Compensation: Employing electronic circuits that monitor temperature and adjust the sensor's output gain accordingly.
    • Software Correction: Characterizing the sensor's temperature-dependent drift and applying algorithms in software to correct readings.
  • Material Selection:
    • Using materials with low coefficients of thermal expansion and stable electrical properties over the expected operating range.
    • Employing hermetic sealing to protect sensitive components from humidity and other environmental factors.
  • Calibration:
    • Regular Calibration: Periodically checking and adjusting the instrument against known standards, often at multiple points across its measurement range to correct for sensitivity.
    • Multi-Point Calibration: Performing calibration at various temperatures or other ambient conditions to create a comprehensive correction map.
  • Environmental Control:
    • Operating instruments within a controlled environment (e.g., temperature-controlled rooms, sealed enclosures) to minimize exposure to varying ambient conditions.
  • Burn-In and Aging Processes:
    • Subjecting instruments to initial stress (e.g., high temperature) before deployment can stabilize their characteristics and reduce early-life drift.
  • Advanced Digital Signal Processing:
    • Implementing sophisticated algorithms that learn and adapt to an instrument's drift patterns over time, applying real-time corrections.

Sensitivity Drift vs. Zero Drift

It's important to distinguish sensitivity drift from zero drift (or offset drift), although both are forms of instrument inaccuracy.

Feature Sensitivity Drift Zero Drift (Offset Drift)
Definition Variation in the instrument's measurement gain or scale Shift in the instrument's output when the input is zero
Impact Error is proportional to the measured value Error is constant, independent of the measured value
Visual Example Slope of calibration curve changes Calibration curve shifts up or down parallel to itself
Correction Re-scaling, gain adjustment, temperature compensation Re-zeroing, offset adjustment, nulling

Both types of drift can occur simultaneously and often require different mitigation strategies to ensure overall measurement accuracy. Effective sensor calibration practices address both.

Understanding and managing sensitivity drift is fundamental for any application requiring precise and reliable measurements under varying environmental conditions. By carefully considering material science, electronic design, and calibration protocols, engineers can minimize its impact and ensure instrument integrity.